Technical data
48
9 Technical data (to DIN 58966)
Low-temperature thermostats
RE 204 RE 206 RE 207 RE 212 RE 220
Operating temperature range °C -10 to 200 -20 to 200 -35 to 200 -30 to 200 -30 to 200
Ambient temperature range °C 5...40
(
#
Section 5.1)
Setting resolution °C 0.01
Indication resolution °C 0.05
Indication accuracy ±0.2 °C
➂
additive re-calibration
Temperature control ± °C 0.01 0.01 0.02 0.02 0.02
Safety features
➀
FL
Additional functions 2-line LCD-display
autom. compressor control autom. compressor control + cooling adjustments
Heater
power
230 V; 50/ 60 Hz
115 V; 60 Hz
100 V; 50/60 Hz
kW 2.25
1.3
1.0
Cooling capacity (eff.)
@ with ethanol at 20 °C
ambient temperature
20°C
0°C
-20°C
-30°C
kW 0.15
0.10
0.05*
----
0.20
0.15
0.05
----
0.30
0.22
0.10
0.06
0.30
0.23
0.13
0.04
0.35
0.25
0.10
0.04
Pump type pressure pump with choice of 5 output steps
Max. discharge pressure
➁
bar 0.4
Max. flow rate
➁
l/min 17
Pump connections mm nipples 13 mm dia.
Max. bath volume l 3 to 4.5 4 to 6 4 to 6 9 to 12 14 to 20
Bath opening (W x D) mm 130x105 150x130 150x130 200x200 300x350
Bath depth mm 160 160 160 200 160
Usable depth mm 140 140 140 180 140
Height top edge of bath mm 363 396 396 441 441
Overall size (W x D x H) mm 180x320x524 200x400x557 200x400x557 250x400x602 350x530x602
Weight kg 21 24 26 30 34
Power
con-
sumption
230 V; 50/ 60Hz
115 V; 60 Hz
100 V; 50/60 Hz
kW 2.5
1.4
1.1
2.6
1.4
1.1
2.7
1.4
1,1
2.7
1.4
1.1
2.8
1.4
1.2
Ref. No. 230 V; 50/60 Hz LCK 0865
230 V; 50 Hz LCK 1862 LCK 1868 LCK 1871 LCK 1874
230 V; 60 Hz LCK 2862 LCK 2868 LCK 2871 LCK 2874
115 V; 60 Hz LCK 4862 LCK 4865 LCK 4868 LCK 4871 LCK 4874
100 V; 50/60 Hz LCK 6862 LCK 6865 LCK 6868 LCK 6871 LCK 6874
*@ -10°C
➀
FL: suitable for flammable and non-flammable liquids
➂
➩
Section 7.2 last item
➁
at pump output step 5
We reserve the right to make technical alterations!