Emptying
Switch off the thermostat, pull out the mains plug!
b) Drain the bath liquid through the drain cock
!
using tubing.
The units are designed for operation with non-flammable and flammable liquids to
EN 61010-2-010! Flammable liquids can be operated up to no more than 25°C
below the firepoint (
Section 5.3.).
Observe the appropriate regulation when disposing of used thermostating liquid.
When connecting an external consumer take care that the level of the bath liquid
does not decrease too much
!
fill in bath liquid if necessary.
Do not drain the thermostating liquid when it is hot or very cold (below 0°C)!