Startin
g
u
p
36
RE 2xx), until the PARA function is reached, then
and
2x
switch to level 2 and move forward with key on the left
The display shows the current bath temperature and the
currently set interface. To alter the setting press the key.
or
Display flashes (approx. 4 s.). During this time start to
set the required interface.
Having chosen the RS 485 interface press the key.
The display shows the current bath temperature, ADR
and the current unit address (e.g.. 15).
Press the key, display flashes (approx. 4 s.)
or
During this time start to set the required unit address.
(possible setting: 000...127).
Forward with key to ”Setting the baud rate” or
back with key to ”Selection of the interface” resp. when
pressing the key twice you return to ”Setting the reset
time...”“
6.4.5.4 Setting the Baud rate/ transfer rate (serial interface)
and
4x
a
nd
3x
Directly from level 0 (setpoint selection), press the key
combination on the left (top one on RE 3xx, bottom one
RE 2xx), until the PARA function is reached, then