EasyManua.ls Logo

Lauda Ecoline RE 207 - Pump Motor Monitoring; Connection Floating Contact Combination Fault 31 N

Default Icon
69 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Startin
g
u
p
42
6.5.3 Pump motor monitoring
In case of pump motor overload or a blockage the heating and the pump are
switched off.
"
""
"

Double signal tone sounds.
The display shows PUMP and the fault triangle is
flashing
Rectify the cause of the fault, e.g. clean the pump or
check the viscosity, then
reset with the key.
If several faults appear simultaneously, they have to be
reset individually
6.5.4 Connection floating contact ”Combination fault” 31 N
(Alarm out) 3-pole locking connector
1 = common, 2 = n.c. (make) , 3 = n.c. (break). When the unit is o.k. 1 and 3 are closed (see ill.). Max
contact rating: 0.2A 24 V. 3-pin plug (
Section 10. Accessories).
Contact operates in case of error at overtemperature protection, low-level
protection, pump monitoring or any other error message.

Table of Contents

Related product manuals