Safety notes
10.08.99
YACE0060
45
Depending on the bath liquid used and the mode of operation it is possible for toxic vapours to be
produced. Ensure appropriate ventilation!
When changing the bath liquid from water to oil, for temperatures above 100 °C, carefully remove all
traces of water, also from tubing and from the external consumer, otherwise
!
danger of burns
through delayed boiling!
Always pull out the mains plug before cleaning, maintenance or moving the thermostat!
Repairs on the control unit and the refrigeration system must be carried out by properly qualified
personnel only
Values for temperature control and indicating accuracy apply under normal conditions according to
DIN 58966. High-frequency electromagnetic fields may under special conditions lead to unfavourable
values. This does not affect the safety