Table 13: Actuating signal
ID Function Unit, resolu-
tion
Command
11 Resolution of controller actuating signal in per mill
– negative value à device is cooling
– positive value à device is heating
[]
IN_PV_06
13 Controller actuating signal in watts
– negative value à device is cooling
– positive value à device is heating
[W] IN_PV_08
Table 14: Cooling
ID Function Unit Command
24 Cooling mode: 0 = o / 1 = on / 2 = automatic [–] IN_SP_02
Table 15: Safety
ID Function Unit Command
35 Timeout communication via interface (1 – 60 seconds; 0 = O) [s] IN_SP_08
73 Status of Safe Mode: 0 = o (inactive) / 1 = on (active) [–] IN_MODE_06
162 Heating output restriction [%] IN_SP_13
166 Heating output restriction in Safe Mode [%] IN_SP_14
168 Pump pressure set point in Safe Mode [bar] IN_SP_15
170 Flow rate set point in Safe Mode, only for device with flow controller [l/min] IN_SP_16
172 Pressure limitation in Safe Mode, only for device with flow controller [bar] IN_SP_17
Table 16: Control parameters
ID Function Unit Command
39 Control parameter Xp [–] IN_PAR_00
41 Control parameter Tn (181 = O) [s] IN_PAR_01
43 Control parameter Tv [s] IN_PAR_02
45 Control parameter Td [s] IN_PAR_03
47 Control parameter KpE [–] IN_PAR_04
49 Control parameter TnE [s] IN_PAR_05
51 Control parameter TvE [s] IN_PAR_06
53 Control parameter TdE [s] IN_PAR_07
55 Correction limitation [K] IN_PAR_09
V6 Integral Process Thermostats and High-Temperature Thermostats 55 / 198