System Commands
I
I
ACQUISITION
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
Response Format
EXAMPLE
SAMPLE_CLOCK, SCLK~
Command/Query
The SAMPLE_CLOCK command allows the user to control the
use of an external time base. The user sets the number of data
points that will be acquired when the oscilloscope is using the ex-
ternal clock.
Sample_CLocK <state> [,<recordlength>]
<state> := {INT,EXT} <recordlength> := {50, 100, 200, 500,
1000, 2000, 5000,
~oooo, 2oooo,
50000)
Note: If <recordlength> is not specified the previous value will not
be modified. (The parameter <recordlength> is initially set to
50000).
Sample_CLocK?
SampleCLocK <state>,<recordlength>
The following command sets the oscilloscope to use the external
clock with 1000 data point records.
CMD$ = "SCLK EXT,1000": CALL IBWRT(SCOPE%,CMD$)