X-Stream Operator’s Manual 
WM-OM-E Rev I  411 
Parameter 
Settings 
When you select acsn from the PRML dialog's With parameter field (DRIVE 
ANALYSIS Æ Measure Æ Parameter Set Æ  PRML) an acsn Pattern Length field 
appears also. 
The pattern length should be set as an integral number of waveform periods, and 
must be at least 50 samples at the DDA’s sample rate. (The sample rate is shown 
in the TimeBase label box.) Because these periods will be correlated with the 
same number of following periods, the pattern length must be no more than half 
the number of full periods available in the sweep. 
Notes on How Settings Affect ACSN Measurement 
Pattern length should be close enough to the signal period (or multiple thereof) 
that it is on the correlation peak. The code measures correlation at 3 positions: the 
length, 6 samples less and 6 samples more. It will walk up a slope until the center 
sample is highest, tighten the spacing until the two side positions have correlation 
above half the correlation at the center position, and eventually perform a 
parabolic fit. But if pattern length misses the peak entirely then this search may not 
work   if the smallest of the three correlations is at the center and the two side 
position correlations are equal, it stops. For a sinusoidal signal, almost any setting 
of pattern length should work.  
One limitation to know about at this stage: the code checks for a "reasonable" 
peak. The correlation at the center position must be > 0.9, or the parameter 
computation is canceled. This is done without the benefit of the parabolic fit, but 
the intent is only to make sure of a reasonable peak. 
The total number of cycles in the acquisition does not affect the ACSN result for a 
stationary signal. Note that horizontal scale and the Max Sample Points entry on 
the TIMEBASE "Smart Memory" dialog interact. As you increase the time of an 
acquisition, the sample rate may decrease to keep the acquisition smaller than 
Max Sample Points. Although ACSN is not affected by horizontal scale, it is 
affected by sample rate. Be sure to keep it high enough. 
For the same signal, ACSN will fall as vertical scale is increased, because the 
signal fills less of the code space of the ADC and is therefore more contaminated 
by quantization noise and any other internal noise. In general, when displayed on 
a channel (or a reset trace), the signal should fill as much of the grid vertically as 
possible, but should never clip against the top or bottom of the grid. If high ACSN 
readings (greater than or equal to 30 dB) are expected, this is especially 
important. 
The pattern length should be kept to the minimum number of cycles that will 
satisfy the 50 samples per pattern requirement. In general, with a somewhat noisy 
signal, as the number of cycles in one pattern length is increased, ACSN is slightly 
reduced. For example, with an 80 MHz 75 mV sine wave with 0.6% (of p-p) rms 
noise, sampled at 1GS/s, acsn result varied with pattern length as follows: