6-20 Maintenance
previously installed will need to be reinstalled. The following steps must be
followed:
• Plug a keyboard into the instrument.
• Insert the System Recovery CD into the CDROM drive.
• Turn on power to the instrument, the system will boot from the CD (if it does not
boot from the CD, the BIOS settings regarding boot order may have been
changed, restart the computer while pressing F2 on the keyboard to get to the
BIOS settings screen. Change the boot order to boot from the CD first).
• After booting the screen should display the X-Stream System Recovery CD
screen.
• Select Continue , press enter
• An additional warning “All C drive data will be lost” appears – press enter for YES
• The software installation progress bars will appear, this loading process will take
app. 5 minutes.
• Remove CD when prompted - press enter
• Wait for the operation to take place (10-15 minutes) and an A:\> prompt to
appear. (If the message “ cannot load the file a:\command.com insert correct disk
and strike a key” appears, cycle power on the scope.
• Remove the CD ROM from the drive and cycle power.
• Insert the Wavemaster Installation CD ROM into the drive.
• When the system boots up, Windows 2000 setup will need to be completed
before the XStreamDSO Software can be installed onto it Accept the license
agreement and press Next.
• You will be prompted for the product key, this can be found on the Windows
License label on the rear panel of the instrument.
• When prompted for Computername, type the serial number located on the rear
panel (e.g. LCRY0601P63247).
• The unit will finish the setup process.
• At network setup press next.
• When LeCroyuser setup comes up, press next (with a blank password). Press
finish.
• Close the add new hardware dialog box.
• Follow directions above for installing new XStreamDSO software.
• Once complete with the XStreamDSO software and driver installation, launch the
DSO application.
• Perform the touch screen calibration procedure in step 6.6.2.
• Installation is complete.