EasyManua.ls Logo

LeCroy WAVERUNNER 6000 SERIES - Specifications; 3.1 Vertical System; 3.2 Horizontal System

Default Icon
325 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications 3-1
3. Specifications
3.1 Vertical System
WaveRunner WaveRunner WaveRunner WaveRunner WaveRunner
6030(A) 6050(A) 6051(A) 6100(A) 6200(A)
Nominal Analog Bandwidth @
50 (-3 dB)
350 MHz 500 MHz 500 MHz 1 GHz 2 GHz
Rise Time (Typical) 1 ns 750 ps 750 ps 400 ps 225 ps
Input Channels 4 4 2 4 4
Bandwidth Limiters 20 MHz; 200 MHz
Input Impedance 1M//<20pF (10 M // 9.5pF using PP007 probe)
Input Coupling 50 : DC, 1M: AC, DC, GND
Maximum Input Voltage, 50 50 : 5 Vrms, 1 M: 250 V max (Peak AC: 10 kHz + DC)
Channel to Channel Isolation >40dB @ <100MHz (>30dB @ full bandwidth)
Vertical Resolution 8 bits; up to 11 with enhanced resolution (ERES)
Sensitivity 50 : 2 mV/div - 1 V/div fully variable; 1 M: 2 mV - 10 V/div fully variable
DC Gain Accuracy ±1.0% of full scale (typical); ±1.5% of full scale 10mV/div (warranted)
Offset Range 50 : ± 400 mV @ 2-4.99 mV/div
± 1 V @ 5-100 mV/div
± 10 V @ 102 mV/div - 1V/div
1 M: ± 500 mV @ 2-4.99 mV/div
± 1 V @ 5-100 mV/div
± 10 V @ 102 mV/div - 1V/div
± 100 V @ 1.02V/div - 10V/div
Offset Accuracy
±(1.5% + 0.5% of offset value + 1 mV) all fixed gain < 2V/div
±(1.5% + 1.0% of offset value + 1 mV) all variable and V/div settings 2V/div
Probing System BNC or Probus®
3.2 Horizontal System
Timebases
Internal timebase common to all input channels; an external clock may be
applied at the auxiliary input
Time/Division Range 20 ps/div - 10 s/div
Math & Zoom Traces 4 independent zoom and 4 math/zoom traces standard
Clock Accuracy 5 ppm @ 25° C ( 10ppm @ 5-40° C)
Time Interval Accuracy Clock Accuracy + Jitter Noise Floor
Sample Rate & Delay Time Accuracy Equal to Clock Accuracy
Trigger & Interpolator Jitter (RMS) 3 ps rms (typical)
Channel to Channel Deskew Range ±9 x T/div setting, 100ms max, each channel
External Sample Clock
DC to 1 GHz; 50 (limited BW in 1M), BNC input, limited to 2 Ch
operation (1 Ch in WR6051), (minimum rise time and amplitude
requirements apply at low frequencies)
Roll Mode User selectable, Available at lower time/div settings

Related product manuals