5-18 Performance Verification WR6K Rev. D
For the low sensitivities: 2 mV through 20 mV/div., connect the test
equipment as shown in Figure 5-1.
DC Power SupplyDMM
20
DB
20
DB
Figure 5-1 : DC 50 Accuracy Equipment Setup for 2 mV - 20 mV/div
For the sensitivities: 50 mV and 100 mV/div, connect the test equipment as
shown in Figure 5-2.
DC Power SupplyDMM
20
DB
Figure 5-2 : DC 50 Accuracy Equipment Setup for 50 mV and 100 mV/div
For the range 200 mV, 500 mV and 1 V/div no attenuator is required, connect the
test equipment as shown in Figure 5-3.
DC Power SupplyDMM
Figure 5-3 : DC 50 Accuracy Equipment Setup for 200 mV, 500 mV and 1 V/div.