5.7.1 Positive 1M DC Accuracy
Procedure
Recall DC accuracy - 1 Mohm 2mv.lss or configure the DSO :
Panel Setups : Recall FROM DEFAULT SETUP
Channels Trace ON Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling : DC
1M on all 4 Channels
Input offset : 0.0 mV on all 4 Channels
Input gain : from 2mV/div to 2 V/div. (see Table 6) on all 4 Ch
C1 Averaging : 5 sweeps
C2 Averaging : 5 sweeps
C3 Averaging : 5 sweeps
C4 Averaging : 5 sweeps
Trigger setup : Edge
Trigger on : Line
Mode : Auto
Time base : 1 µsec/div.
Channel use : 4
Change parameters
P1 : Measure mean of C1
P2 : Measure mean of C2
P3 : Measure mean of C3
P4 : Measure mean of C4
For the low sensitivities: 2 mV through 20 mV/div., connect the test
equipment as shown in Figure 5-4.
DC Power SupplyDMM
20
DB
20
DB
50
Figure 5-4 : DC 1M Accuracy Equipment Setup for 2 mV - 20 mV/div
For the sensitivities: 50 mV and 100 mV/div, connect the test equipment as
shown in Figure 5-5.
WR6K Rev.D Performance Verification 5-21