5-28 Performance Verification WR6K Rev. D
5.8.3 Positive 1M Offset Accuracy
Procedure
Recall Offset - 1 Mohm Positive.lss or configure the DSO:
Panel Setups : Recall FROM DEFAULT SETUP
Channels Trace ON Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling : DC 1M on all 4 Channels
Input gain : 50mV/div on all 4 Channels
Input offset : +0.750 Volt on all 4 Channels
C1 Averaging : 5 sweeps
C2 Averaging : 5 sweeps
C3 Averaging : 5 sweeps
C4 Averaging : 5 sweeps
Trigger setup : Edge
Trigger on : Line
Mode : Auto
Time base : 50 ηsec/div.
Channel use : 4
Statistics : on
Change parameters
P1 : Measure mean of C1
P2 : Measure mean of C2
P3 : Measure mean of C3
P4 : Measure mean of C4
Connect the test equipment as shown in Figure 5-7.
Set the output of the external DC voltage reference source until the DVM
measures 0.750 Volt.
1) Verify that the displayed trace A : Average (1) is on the screen, near the center
horizontal graticule line. If the trace is not visible, modify the DC voltage
reference source output until the trace is within ± 2 divisions of center.
2) Connect the DMM and record the voltage reading in Table 11, column DMM.
3) Disconnect the DMM from the BNC T connector.
4) Press Clear Sweeps
5) After 100 sweeps, Read off the DSO Mean parameter voltage, and record the
measurement in Table 11, column Mean.