EasyManua.ls Logo

LeCroy Waverunner2 LT584M

Default Icon
155 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rev. D 10
of
12
LT584 & LT374/372 & LT354 & LT264/262 Test Record
Trigger
Level
Trigger
Slope
Channel 1 Channel 2 Channel 3 Channel 4 Lower
Limit
Upper
Limit
mV
Measured
DC
Trigger
Level (1)
mV
Measured
DC
Trigger
Level (2)
mV
Measured
DC
Trigger
Level (3)
mV
Measured
DC
Trigger
Level (4)
mV mV mV
0Pos
30
+30
0Neg
30
+30
+300 Pos +270 +330
+300 Neg +270 +330
300
Pos
270
330
300
Neg
270
330
Table 10: Channel DC Trigger Test Record
Trigger
Slope
External
Trigger
Level
External
DC
External
Limits
External/10
Trigger
Level
External/10
DC
External/10
Limits
mV
Measured
DC Trigger
Level (Ext)
mV
Lower
MV
Upper
mV V
Measured
DC Trigger
Level (Ext10)
V
Lower
V
Upper
V
Pos 0
50
+50 0
0.5
+0.5
Neg 0
50
+50 0
0.5
+0.5
Pos +300 +250 +350 +3 +2.5 +3.5
Neg +300 +250 +350 +3 +2.5 +3.5
Pos
300
250
350
3
2.5
3.5
Neg
300
250
350
3
2.5
3.5
Table 11: External & Ext/10 DC Trigger Test Record