EasyManua.ls Logo

LeCroy Waverunner2 LT584M

Default Icon
155 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rev. D 11
of
12
LT584 & LT374/372 & LT354 & LT264/262 Test Record
Smart Trigger
Pulse Width
ns
Generator
Frequency
MHz
Width
<
Width
>
Triggered Pass
< 10 100 On Off Yes
< 10 100 Off On No
> 10 40 Off On Yes
> 10 40 On Off No
< 100 10 On Off Yes
< 100 10 Off On No
> 100 4 Off On Yes
> 100 4 On Off No
Table 12: Smart Trigger Test Record
Generator
Frequency
MHz
Post Trigger
Delay
Msec
Delay (A )
µ
s
Delay (1)
msec
Difference
delay(A)
delay(1)+50msec
Lower
Limit
µ
µµ
µ
sec
Upper
Limit
µ
µµ
µ
sec
0.1000 50.0 -0.5 +0.5
Table 13: Time Base Test Record