63
Leica IP C
Cleaning and Maintenance 6
Fig. 51
1
Fig. 52
• Drying station chute( "Fig. 52")
The swivelling flap must be open
( "Fig. 51-1").
The chute must be clean ( "Fig. 52").
Caution
Sensitive electronics components are
located in this area.
Use no liquid in this area!
• Close the flap after cleaning and before using
the instrument ( "Fig. 50").
Outer surfaces
• Clean the outer surfaces (including those of the automated cassette unload station) with a mild
detergent and subsequently dry with a slightly moistened cloth.
• Do not use any solvents for cleaning the outer surfaces and the lid!
Automated unload station
• Remove the unload trays; with a brush, remove dust and debris from guides and ejector.
• The trays themselves can be cleaned with a household cleaner.
• Do not use any solvents to clean the trays!
• Prior to reinserting them into the instrument, the trays must be completely dry.