76
Version 2.0, Revision I
Troubleshooting
7
Display
code
Source of error Troubleshooting solution
Cassette not correctly clamped in cassette
carrier.
Cassette did leave the feeding chute but
did not reach the cassette carrier.
Remove the cassette from the cassette
carrier.
Cassette did not leave the cassette carrier
or was still located in the cassette carrier
during initialization.
Remove the cassette from the cassette
carrier.
The ink print head is getting too hot.
Ambient temperature too high or
electronics defective.
Switch instrument off and let it cool off.
The instrument remains disabled until
the print head has cooled down to a
temperature value within the permissible
range.
Check ambient temperature.
No or incorrect voltage at ink print head. Call Leica Technical Service.
• Cover ( "Fig. 2-5") not closed properly.
Safety switch triggered.
• Flash standby state not reached within
the prescribed time. Charging electronics
defective.
• Ensure that the cover is not blocked,
e.g. by the manual unload station. Close
cover completely.
• Call Leica Technical Service.
No flash or flash duration too short.
Flash bulb dirty or defective.
Check whether flash bulb works – to do
so, observe the stray light on the cover.
Never open the cover to check whether a
flash is triggered!
No flash --> install a new flash bulb
( p. 32 – 4.6 Installing/exchanging the
flash bulb).
Unload station mechanically blocked. Remove the cause of the blockage.
Heater fan not running or running too
slowly.
Call Leica Technical Service.
No cassette is in the drying module for
flashing. Cassette left the cassette carrier,
but did not reach the drying module.
Remove cassette.
Cassette not successfully ejected from
drying module.
Remove cassette.