77
Leica IP C
Troubleshooting 7
Display
code
Source of error Troubleshooting solution
Cassette in the drying module during
initialization or prior to a print job being
carried out.
Remove the cause of the blockage.
The flap of the drying module either does
not close or does not open. The flap is
blocked (e.g. by a cassette).
Remove the cause of the blockage.
Lifting table end sensor does not switch. Call Leica Technical Service.
Lifting table position sensor does not
switch during positioning.
Call Leica Technical Service.
Flash power supply: Overtemperature for
more than 10 minutes.
Call Leica Technical Service.
Faulty control data received (program bug).
Settings for the serial interface incorrect or
the instrument configuration conflicts with
the PC configuration.
Carry out a RESET on the printer.
Check cable connection to the PC.
Check configuration of serial port of the
PC and reboot the PC.
Transmitted data do not contain
confirmation of receipt or data
transmission was not confirmed by the PC.
Follow the same procedure as for
ERROR 60.
Try another printer cable.
Print image exceeding the vertical limit. Error caused by application software.
Print image exceeding the horizontal limit. Error caused by application software.
The CRC test of the EEPROM returned an
error when the instrument was switched
on.
Call Leica Technical Service.
Internal firmware error or defective
controller.
Call Leica Technical Service.
Mechanical problem makes ejection of
cassettes from magazine difficult.
Check ejection mechanism. Remove
foreign bodies, then clean with brush.

Other manuals for Leica IP C