Loading...
The total magnification
of
the microscope is given
by:
rep
rod
uc
ti
on
ratio
of
the ob
jec
ti
ve x eyepiece magni
fi
ca
ti
on
(x
tube factor).
Example:
25/0
.5
0 objec
ti
ve
IOx/18
eye
pi
ece
tube fact or Ix
To
tal magnificatio n:
25
x
10
x 1 =250x
Scre
wing
in
the objectives
Screw the o
bj
ec
ti
ves into the nosepi
ece
(Fig.
4)
in
such a way
that stepwise magnifica
ti
on
chan
ge
is
possible (e.g. in the order
4,
10
, 40, etc).
All Leitz microscope obje
ct
ives from a magnification
of
2.5x
and calculated for a tu be length of160 mm
ca
n be used on th e
Lei tz L
AB
ORLUX
S.
Please no te
in
th
is
cootext the field
iUu
-
mjnation capac
it
y
of
the
condense
r
s.
Microscope o
bj
ec
ti
ves calcul ated for
17
0 mm tube length can
be used from 1
6x
magnifica
ti
on.
Fig.
4
Sc
r
ew
ing
in
th
e objectives
I