Technical update
Technical update ICAR approved Shuttle XY
1 Introduction
The milk recording device Shuttle XY is tested for a period of time resulting in ICAR approval. The approval is applicable for the Shuttle
XY on all Astronaut milking robots. Some modifications are required to comply to the ICAR standard. Resulting in better mixing of the
milk in the milk jar, reducing cleaning water remaining and ensuring minimal carry-over during sampling.
This document will describe required modifications to comply to ICAR and will be a guideline for correct sampling.
2 ICAR update matrix
Required modifications can be found in the update matrix and will be explained below.
Modifications are required to comply to ICAR
Y= needs update
N/A = Not applicable, no update is required.
9.1065.0074.2
(100m)
Silicone tube
Inner diameter to 4mm
Max. length of sample tube
1meter.
Y:
SN <3170637 (4/10/’13)
N/A:
for newer models
5.1004.1615.0
Connection
nipple
Y:
SN <3170637 (4/10/’13)
N/A:
for newer models
5.1004.0226.0
Milk manifold /
Pre-milk
manifold
bigger drain holes, 10.5mm
(inside)
bigger drain pipe inside to 12mm
Y:
SN <3157694 (5/3/’13)
N/A:
For newer models
bigger inner diameter of PRE
drain 6-folded tube 14mm
Y:
SN <3161129 (19/4/’13)
N/A:
For newer models
5.1001.6659.0
Nipple (2x)
Bigger connector pipe of drain
tube. Inner diameter increased,
13mm
Y:
SN <3161129 (19/4/’13)
N/A:
For newer models
9.1185.0013.3
(30m)
Tube (2x)
Bigger inner diameter (14mm)
drain tube at separator
Y:
SN <3161129 (19/4/’13)
N/A:
For newer models
Robot Software
4.1 or newer
Version with improved sampling
process A3N and A4 and longer
addition pumping time A2 for
data communication with
Shuttle XY