15
4 Operation
4.1 General Timing Specifications
4.1.1 Startup Time
During start-up (after power-on) the flow measurement is not performed. Start-up time is about 8 seconds.
4.1.2 Thermal Stability Time
To guarantee specified accuracy thermal stability of sensor must be reached. Therefore, it is recommended to
wait 30 minutes after power-on to get stable flow measurements.
4.1.3 Zero Adjustment
The duration to perform a zero adjustment is < 30 seconds. Following options are available for zeroing:
1. Activating by Levitronix
®
Service Software.
2. Activating by the user panel in the system menu functions.
3. Activation by digital input (default configuration).
4.2 Standalone Operation with User Panel
Only the basic functions of the user panel are described here. For more detailed description consult the
standard firmware specification of the LUI-B.1-02.
4.2.1 Basic Operation
With the user panel LUI-B.1 the flowmeter can be operated and monitored manually according to Figure 11..
After power on the display shows the actual flow. By pressing the Up or Down button the actual calibration set
can be viewed. By pressing the Up or Down button again the user returns to the flow monitoring display.
Act. Flow [ml/min]
1000
Figure 11: User panel for manual operation
4.2.2 System Menu Functions
By pressing the On/Off button for more than 2 seconds the System Menu can be reached with the following
basic functions:
SETTINGS
ZERO ADJUSTMENT
INFO
1
a. Monitor the flow.
b. Selection of calibration set.
c. Volume counter.
2
a. Zero adjustment of flow meter.
3
a. Flow and full scale.
b. Serial number and sensor size.
c. Firmware and hardware revision.
d. Status.
Figure 12: System Menu and its basic functions
In the System Menu and its sub menus, the menu tabs are browsed using the Up and Down buttons. Items
are selected with the On/Off button, and values are modified using the Up and Down buttons. The Menu can
be left by pressing the On/Off button for more than 2 seconds.
SENSOR-TYPE
LFSC-i06X
CFlex, 37°C
Menu
Down
Display
Actual Calibration
Set
Up

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Levitronix LEVIFLOW LFSC-i16X and is the answer not in the manual?

Levitronix LEVIFLOW LFSC-i16X Specifications

General IconGeneral
BrandLevitronix
ModelLEVIFLOW LFSC-i16X
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals