User Manual for PuraLev
®
i30SU
www.levitronix.com
10
2.4 Pressure-Flow Curves
Figure 11: Pressure/flow rate curves of DCP-30 pump head
2.5 Maximum Static Pressure for Pump Heads
Figure 12: Specification for maximum static pressure of pump head DCP-30
Note 1: Pressure limits only valid for pump head mounted on motor and do not account for tubing limits attached to the fittings.
Note 2: The above chart is also applicable to the DCP-30.2-SF1-G25 with sterile fittings.
0
2
4
6
8
10
12
14
16
0.0 0.5 1.0 1.5 2.0
0.0
0.2
0.4
0.6
0.8
1.0
0 1 2 3 4 5 6 7 8
[psi]
[gallons / min]
[bar]
[liters / min]
6000 rpm
13000 rpm
12000 rpm
11000 rpm
10000 rpm
8000 rpm
Specific gravity = 1g/cm
3
Viscosity = 0.7cP
Liquid Temp.: 40C
9000 rpm
7000 rpm
5000 rpm
0
10
20
30
40
50
60
70
80
77 87 97 107 117 127 137
0
1
2
3
4
5
6
25 30 35 40 45 50 55 60
[psi]
F]
[bar]
Liquid Temperature in [°C]
Max. Static Pressure for DCP-30
Safety Factor = 1.3 (according to EN12162)
Note: Burst pressure is significantly higher.
Ambient Temp = 25 C
Note: Pressure limit only valid for pumphead mounted on motor. Values
only valid after Gamma radiation doses of < 40 kGy.