4.3 Electrical Installation of OEM Model
4.3.1 Overview Wire Designation of Driver Cable
Wire
Name
Wire Color
UL Cable
Description
Standard
Designation
Typical Levels
Note
P+
red
+ 24 VDC
Supply
Voltage: 24 VDC ±10%
P- to be connected to earth.
P-
black
Power Input
Ground / Earth
Ain1
violet
Analog Input 1
(Current Input)
Actual Process
Value
4..20 mA = 0..100%
The configuration of the analog inputs can be changed
with Levitronix
®
Service Software
Ain2
grey-
pink
Analog Input 2
(Voltage Input)
Reference
Value
0..10 V = 0..16000 rpm (Speed Mode)
-> Speed limit = 13000 rpm
-> Cut-off (min.) speed = 0 rpm
0..10 V = 0..100% (Process Mode)
The configuration of the analog inputs can be
changed with Levitronix
®
Service Softw. 2.0
Note: Max. input voltage of 11 V shall not be
exceeded.
Ain_GND
blue
Analog Input
Ground
--
--
Reference for analog inputs.
Din1
pink
Digital Input 1
Enable
(Reset)
5-24 V active
0 V not active
The “Enable” signal switches the pump system on
and off.
Resets pump from error state with 300-700ms pulse.
Din2
grey
Digital Input 2
Process Mode
5-24 V active
0 V not active
Switch between process mode and speed mode. In
process mode the closed loop PI controller can be
configured with Levitronix
®
Service Software
Din_COM
yellow
Common
Digital Input
--
--
--
Aout1
blue-
red
Analog Output
(Voltage Output)
Actual Speed
Actual Process
Value
0..10 V = 0..16000 rpm (Speed Mode)
0..10 V = 0..100% (Process Mode)
Direct connection, no protection. Galvanic isolation
on user side is required. AGND is reference.
The configuration of the analog output can be changed
with Levitronix
®
Service Software.
Dout1
brown
Digital Output 1
Status
Closed circuit active, system on
Open circuit not active, system off
This signal indicates the state of the pump system.
The configuration of the digital outputs can be
changed with Levitronix
®
Service Software.
Dout2
white
Digital Output 2
Error
Closed circuit not active, system on
Open circuit active, system off
When active, the system drives the impeller to zero
rpm and shuts down. With a reset pulse (see
Enable” description) the system can be re-
initialized.
The configuration of the digital outputs can be changed
with Levitronix
®
Service Software.
GND
white-green
Analog Ground
--
--
Reference for Aout1, Dout1 and Dout2
RS485+
brown-
green
RS485 +
Field Bus
Modbus protocol
Termination resistors recommended
RS485-
white-
yellow
RS485 -
Internal
yellow-
brown
Internal
Do not connect.
--
For internal usage.
Internal
white-grey
Internal
Do not connect.
--
For internal usage.
Shield
shield
wire
Shielding
Shielding
To be connected to earth
(see wire P-, black)
Table 13: Signals of the driver cable with designation for standard firmware
Note 1: For other configurations of PLC Inputs and Outputs refer to alternate firmware documentation
Note 2: Configurations can be done with Levitronix
®
Service Software.
Note 3: Power wires (P+, P-) have cross section 1.5 mm
2
and all others 0.14 mm
2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Levitronix PuraLev i30SU and is the answer not in the manual?

Levitronix PuraLev i30SU Specifications

General IconGeneral
BrandLevitronix
ModelPuraLev i30SU
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals