4.3.2 Settings converter
Menu item Description Adjustable value / option Ac-
cess
min. max. stan- Unit
value value dard
Option Relay 1 Relay with change-over Threshold bearing
(Relay 6) contact; the operator can temperature reached °C r/w
select one of the functions Threshold motor
described current reached A
Threshold frequency
the thresholds are adjusted reached Hz
in the following menu items No cooling water -
no purge gas -
TMS temperature o.k. -
Vent -
Start command applied -
Power supply o.k. -
Pump standstill -
Option Relay 2 Start command applied -
(Relay 7)
Option Relais 3 Power supply o.k. -
(Relay 8)
Option Relais 4 Pump standstill -
(Relay 9)
Threshold Threshold bearing temp. 0 100 60 °C r/w
for option relay
Threshold Threshold motor current 0 20 10 A r/w
for option relay
Threshold Threshold frequency 0 480 300 Hz r/w
for option relay
Analog Output Analog output 0...10V; Motor current;
the operator can select 20 A =10Vxscale factor V r/w
one of the functions Act. frequency ;
described 500 Hz = 10Vxscale factorV
Bearing temp.;
100 °C = 10Vxscale factor V
Actual PVW13;
100% = 10Vxscale factor V
Actual PVW24;
100% = 10Vxscale factor V
Actual PZ12;
100% = 10Vxscale factor V
Scale Factor Scale factor for the analog output 0 2 1 - r/w
Analog Input 2 Analog input 0...10V ; No function - r/w
optionally frequency setpoint
via analog input 2 Function frequency
(10V = max. frequency setpoint) setpoint -
Start Delay Waiting time between start
command and acceleration 0 1200 0 s r/w
Display Display language English - r/w
German -
Bus Address Bus adress for the converter by
operation via 0 31 0 - r/w
serial interface RS232/485
German
Start Command
Enter
Threshold
60 °C
Enter
Threshold
10.0 A
Enter
Threshold
300 Hz
Enter
Actual PZ12
Actual PVW24
Actual PVW13
Bearing Temp.
Actual Frequency
Analog Output
Motor Current
Enter
Scale factor
1.0 x
Enter
Freq. Setpoint
Analog Input 2
No Function
Enter
Start Delay
0 s
Enter
Display
English
Enter
Bus Address
0
Enter
Enter
Option Relay 3
Power Supply OK
Enter
Option Relay 4
Pump Standstill
Enter
Option Relay 2
Start Command
The setttings of the option relays
2 to 4 can be changed similar to
option relay 1 but they cannot be
stored in the pump’s data storage
device. After each power off the
option relays 2 to 4 are in the
default state.
Vent
TMS OK
No Purge Gas
No Cooling Water
Frequency
Set Converter
Enter
Motor Current
Option Relay 1
Bearing Temp.