GA05141_0702 - 09/2004
19
Description
Technical data (continued)
Purge gas
see Section 2.6
MAG.DRIVE
digital
Voltage range 200 - 240 V +10% -15%
Line supply frequency 50 / 60 Hz
Load
Stand-by approx. 100 W
Maximum heated pumps 1800 W
Maximum non-heated pumps 1100 W
Max. voltage motor 60 V
Maximum pump current 15 A rms
Internal main fuse 16 A
Rated braking capacity 200 kA
System fuse L or G characteristic
Maximum frequency 600 Hz
Load capability, relay output 42 V, 1 A
Temperature during operation 0-45 °C
Storage temperature - 10 °C to + 60 °C
Relative air humidity 95% (non condensing)
Overvoltage category II
Contamination level in accordance with EN 61010 2
Weight 10 kg
Degree of protection (EN 60529) IP20