Maintenance
29GA01601_002_C1 - 03/2018 - © Leybold
5.3 Monitoring the Oil Level
During operation of the TRIVAC the oil level must always remain between
marks (6/2) and (6/3) on the oil-level glass. The amount of oil must be che-
cked and topped up as required.
Check and top up oil only after having shut down the pump first.
Please note the safety information given in Section 0.3 and 0.4.
Pumps which have not directly been delivered from Leybold, e.g. in systems,
may have been filled with oils different from LVO 310. In this case a refill with
LVO 310 may cause problems. Therefore check before the refill the kind of oil
in the pump and fill in only suitable oil or change the oil.