Maintenance
30 GA01601_002_C1 - 03/2018 - © Leybold
2
1
3
4
Fig. 6 Oil change
1 Oil fill plug
2 Oil level mark “maximum”
3 Oil level mark “minimum”
4 Oil drain plug
5.4 Oil Change
Pump and operating agents may be contaminated
Hazardous substances may escape from the pump and the oil. If there is
the danger that the operating agent may present a hazard in any way due
to decomposition of the oil, or because of the media which have been
pumped, you must determine the kind of hazard and ensure that all neces-
sary safety precautions are taken use gloves, face protection or a respira-
tor, for example.
Observe Safety Informations 0.3 und 0.4.
Change the oil while the pump is cold so as to avoid releasing adsorbed
gases. If releasing adsorbed gases is no problem, change the oil after the
pump has been switched off and while the pump is still warm.
Risk of suffering burns
With the pump warm from operation, both pump and oil can get so hot
that there is the risk of suffering burns. If required, wear gloves.
Check and top up oil only after having shut down the pump first.
When disposing of waste oil, observe the applicable environment pro-
tection regulations!
When changing the oil use the same type of oil which was previously in the
pump. If you want to change the type of oil entirely please consult us first.
DANGER
NOTICE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LEYBOLD TRIVAC D 2,5 E-LD and is the answer not in the manual?

LEYBOLD TRIVAC D 2,5 E-LD Specifications

General IconGeneral
BrandLEYBOLD
ModelTRIVAC D 2, 5 E-LD
CategoryWater Pump
LanguageEnglish

Related product manuals