in the Settings Window, Flow Module tab (see Controlling Flow with the Windows®
Software on the next page).
Power Save – After about 7 minutes of control panel inactivity, the 7200-101 con-
trol panel will enter a power save mode, in which almost all of the display LEDs
will turn off, leaving only the Power Save LED illuminated. Press any other button
or turn the Adjust knob to take the control panel out of power save mode.
Power – Illuminates when power is applied to the instrument. This LED will
always be on when power is applied.
In Memory – The In Memory LED indicates when the seven segment display
matches the parameter stored in memory. The two parameters that can be stored
are the Flow setting and the Address.
Whenever the Set Flow LED is selected, the Flow setting is recalled from memory
and entered on the display, and the In Memory LED will be lit. If the Adjust knob
is turned, the In Memory LED will remain off until the Store In Memory button is
pressed or the Adjust knob is turned until the displayed setting once again matches
the stored setting. This same behavior applies the Address LED.
Store In Memory – The 7200-101 has some built in memory for retaining the
desired Set Flow rate and Address. These values are retained in memory after the
7200-101 is powered off. When the desired flow rate is displayed during adjust-
ment, press the Store In Memory button to save to memory. When the displayed
Set Flow setting matches that which is stored in memory, the In Memory LED
(above) will illuminate. When the display Address matches the address stored in
memory, the In Memory LED will be lit.
Flow ON/OFF – Turns the blower motor drive on or off.
Menu button – Toggles the display between the five menu selections indicated by
the LEDs beneath the display.
Section 1. Overview of the 7200-101 Flow Module
1-9Controlling Flow with Manual Controls