EasyManua.ls Logo

LI-COR LI-6400 - Cant Achieve Low Values; Cant Achieve High Values

LI-COR LI-6400
846 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
6400-01 CO2 Mixer Problems
20-26 Using the LI-6400
20
doing the mixer calibration (page 18-20), to improve the mixerÕs Òfirst guess-
esÓ.
CanÕt Achieve Low Values
Typically, the minimum value achieved by the mixer is 30 to 50 µmol mol
-1
.
If your minimum value is considerably above that, then here are some things
to check:
1 WhatÕs your maximum flow rate?
In new measurements mode, press 2, then f2, then F, then enter 1000, to re-
quest a fixed flow of 1000 µmol s
-1
. Then monitor flow rate (display line b).
If the value is considerably lower than 700 µmol s
-1
, turn the mixer off. If that
dramatically increases the flow rate, then the ÒproblemÓ is that the mixer cal-
ibration has specified a reduced flow rate to achieve higher CO
2
concentra-
tions, with the side affect of increasing the minimum concentration. Redo the
mixer calibration (Calibrating the CO2 Mixer on page 18-20).
2 Is the soda lime on full scrub and is it good?
Turn off the mixer, and verify that the reference CO
2
goes to zero. If it
doesnÕt, there is a problem with the soda lime, or else the IRGA needs to be
zeroed.
3 Bad valve?
See The Bad Valve Saga on page 20-28, to see if this could be the problem.
4 Valve installed OK?
If the valve discussed in The Bad Valve Saga below has been replaced, but
is installed backwards, the only CO
2
level that can be achieved will be very
high, typically above 2000 µmol mol
-1
.
CanÕt Achieve High Values
The CO
2
mixer should be able to achieve an upper limit in excess of 2000
µmol mol
-1
. 2200 is a typical value. If you canÕt come close to that, or if it
takes a long time to get there, check for the following:
1 Flow from the source
See Step 4, Check the regulator for flow on page 20-23.
2 Calibration problem?
Change to C mode (constant control signal), and set a 5000 mV target. If that
fixes the problem, then re-do the mixer calibration.

Table of Contents