Lighthouse ApexRemote Operating Manual
3-6 248083447-1 Rev 1
Date Settings:
Figure 3-7 Instrument Date and Time Settings
Note: Make sure to
update the instrument
settings before the
SmartPort cable is
disconnected from the
ApexRemote.
All ApexRemote products can have their date and time settings
changed in this screen. Refer to Figure 3-7 to change the Instrument
Date / Time by using the Set to System Time. Corrections to errors
can be reversed by clicking Restore Values. Make sure that desired
changes are followed by clicking the Update Instrument button to
save the changes to the instrument memory. If the instrument is not
updated before the SmartPort cable is disconnected, the changes will
be lost.
Factory
Standard
Settings
ApexRemote PoE Model:
The ApexRemote PoE standard network settings are set to the same
values for all ApexRemotes shipped. These values are 10.10.x.xxx
for IP, 255.255.0.0 for netmask and 0.0.0.0 as Default Gateway setting.
To view the ApexRemote web page, type its IP address into a web
browser on a computer in the same network subnet.
The IT group should provide IP, Mask and Default Gateway values that
meet the local network needs. The ApexRemote must be changed to
meet these requirements. Each device will require the IP to be changed
to a new IP, as shown in Figure 3-8, to prevent address conflicts.
This button loads
and time values
computer date the
This button saves
the changes to the
memory
into program
memory
instrument