Lighthouse ApexRemote Operating Manual
5-2 248083447-1 Rev 1
8. After the instrument has met the requirement of the Purge Count
test, return the instrument to its normal location and operating
status.
Fault Isolation If the instrument does not pass the Purge Count test, perform the
following procedure:
1. Verify that the Inlet and Outlet barbs are finger-tight - do NOT
over tighten.
2. Check the data over the last 6 five-minute sample times.
3. If sporadic counts are occurring over all channels, the unit may
still have particles inside it. Allow the unit to sample overnight
with the purge filter attached before retesting it. If the counts are
still high after the overnight purge, call Lighthouse Technical
Support for assistance.
4. If the data shows consistent counts in the smallest channel only,
the instrument may have electrical problems and may need to be
returned to Lighthouse. Call a Lighthouse Service Representative
for assistance.
Instrument
Service
Report
The Instrument Service Report screen shows ApexRemote-specific
and diagnostics information info. The displayed information can be
saved to a pdf file by clicking the "save report" button (see Figure 5-1).
Figure 5-1 Instrument Service Report Screen

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lighthouse Worldwide Solutions ApexR5 and is the answer not in the manual?

Lighthouse Worldwide Solutions ApexR5 Specifications

General IconGeneral
BrandLighthouse Worldwide Solutions
ModelApexR5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals