Lighthouse ApexRemote Operating Manual
A-6 248083447-1 Rev 1
Device Options
If Bit 0 of Register 40050 is set, it indicates that the instrument is
capable of Fast Download.
Device Status
The Device Status registers (40003 and 40057) display the current
status of the device (Table A-5). Additional status bits are shown in
40056 (Table A-6).
.
40074 signed integer ApexRemote:
Last Calibration Date [high]. Indicates when instrument was
last calibrated. This number can be generated by the ANSI C/
C++ time() function.
40075 signed integer ApexRemote:
Last Calibration Date [low]
Table A-3 Sensor Settings Registers
Register Data Type Description
Table A-4: Device Options
Bits Description
5 ApexRemote:
Software controlled RGB LED (1=Enabled, 0=Disabled)
6 ApexRemote:
Location Bracket (1=Enabled, 0=Disabled)
Table A-5 Device Status Status (40003/40057)
Bit Description
0 RUNNING: Set when a start command is executed via
Command 11 (instrument start) or through the user
interface. The flag will remained set until a stop
command is executed.
1 SAMPLING: This is set only when the instrument is
actually sampling data that is to be recorded. Caution
must be used in sending a command during this time that
may invalidate current sample.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lighthouse Worldwide Solutions ApexR5 and is the answer not in the manual?

Lighthouse Worldwide Solutions ApexR5 Specifications

General IconGeneral
BrandLighthouse Worldwide Solutions
ModelApexR5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals