EasyManuals Logo

Lighthouse Worldwide Solutions ApexR5 User Manual

Default Icon
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
ApexRemote MODBUS Register Map v1.50
248083447-1 Rev 1 A-3
40021 ASCII string Model Name char[12], char [13]
40022 ASCII string Model Name char[14], char [15]
40023 unsigned integer Flow Rate. See registers 40041-40042 for flow rate units.
Liquid Particle Counters and Samplers:
Value equals flow rate. For example: 100d=100
All Other Instruments:
Divide by 100 to get flow rate. For example: 100d=1.00
40024 signed integer Record Count. Total number of records stored in the counter
40025 signed integer Record Index. Zero based index to data in 3xxxx register
series. Must be lower than the record count (register 40024).
Set this index to expose a counters record in the 3xxxx
registers. Set to -1 to retrieve last record stored in the counter.
40026 unsigned integer Location number
Particle Counters
: Specifies location of Particle Counter where
data was recorded. Must be 1 to 200 (maps to location names
associated with registers 40200 - 40999).
40027 signed integer Real Time Clock (RTC) [high]. Displays instrument’s real-time
clock. Works in conjunction with 40028. Displays date and
time, in number of seconds since midnight, 1/1/1970. Can be
generated by ANSI C/C++ time() function.
40028 signed integer Real Time Clock [low]
40029 unsigned integer Initial Delay [high]. Works in conjunction with 40030.
Number of seconds to wait before starting the first sample. Max
value is 359,999, which equals 99h 59m 59s.
40030 unsigned integer Initial Delay [low]
40031 unsigned integer Hold Time [high]. Works in conjunction with 40032. Number
of seconds to wait between sample periods. Max value is
359,999, which equals 99h 59m 59s
40032 unsigned integer Hold Time [low]
40033 unsigned integer Sample Time [high]. Works in conjunction with 40034.
Number of seconds to sample. Max value is 86,399, which
equals 23h 59m 59s.
40034 unsigned integer Sample Time [low]
Table A-3 Sensor Settings Registers
Register Data Type Description

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lighthouse Worldwide Solutions ApexR5 and is the answer not in the manual?

Lighthouse Worldwide Solutions ApexR5 Specifications

General IconGeneral
BrandLighthouse Worldwide Solutions
ModelApexR5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals