ApexRemote MODBUS Register Map v1.50
248083447-1 Rev 1 A-15
2 Particle Overflow Status
0 = No overflow 1 = Overflow occurred.
3 Instrument Service Status
0 = Instrument is working correctly.
1 = Service light is on. Instrument malfunction detected.
4 Threshold High Status
0 = Threshold not exceeded. 1 = Threshold exceeded
5 Threshold Low Status
0 = Threshold not exceeded. 1 = Threshold exceeded.
6 Instrument Sampler Status
0 = Nominal Operation. 1 = Sampler Error.
7 ApexRemote:
0 = LASER power in spec. 1 = LASER power out of
spec.
8 ApexRemote:
LASER current status
0 = LASER current in spec. 1 = LASER power out of
spec.
9 ApexRemote:
LASER supply status
0 = LASER supply in spec. 1 = LASER supply out of
spec.
10 ApexRemote:
LASER life status
0 = LASER life in spec. 1 = LASER life out of spec.
11 ApexRemote:
No flow status
0 = There is flow in the unit. 1 = no flow in unit.
12 ApexRemote:
Photoamp supply status
0 = Photoamp supply in spec. 1 = Photoamp supply out
of spec.
13 ApexRemote:
Background status
0 = Photoamp background good. 1 = Photoamp back-
ground out of spec.
Table A-14: Instrument Current Status
Val ue
Action