ApexRemote MODBUS Register Map v1.50
248083447-1 Rev 1 A-15
2 Particle Overflow Status
0 = No overflow 1 = Overflow occurred.
3 Instrument Service Status
0 = Instrument is working correctly.
1 = Service light is on. Instrument malfunction detected.
4 Threshold High Status
0 = Threshold not exceeded. 1 = Threshold exceeded
5 Threshold Low Status
0 = Threshold not exceeded. 1 = Threshold exceeded.
6 Instrument Sampler Status
0 = Nominal Operation. 1 = Sampler Error.
7 ApexRemote:
0 = LASER power in spec. 1 = LASER power out of
spec.
8 ApexRemote:
LASER current status
0 = LASER current in spec. 1 = LASER power out of
spec.
9 ApexRemote:
LASER supply status
0 = LASER supply in spec. 1 = LASER supply out of
spec.
10 ApexRemote:
LASER life status
0 = LASER life in spec. 1 = LASER life out of spec.
11 ApexRemote:
No flow status
0 = There is flow in the unit. 1 = no flow in unit.
12 ApexRemote:
Photoamp supply status
0 = Photoamp supply in spec. 1 = Photoamp supply out
of spec.
13 ApexRemote:
Background status
0 = Photoamp background good. 1 = Photoamp back-
ground out of spec.
Table A-14: Instrument Current Status
Val ue
Action

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lighthouse Worldwide Solutions ApexR5 and is the answer not in the manual?

Lighthouse Worldwide Solutions ApexR5 Specifications

General IconGeneral
BrandLighthouse Worldwide Solutions
ModelApexR5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals