Lighthouse REMOTE-4PN Series Operating Manual
3-8 248083317-1 Rev 4
16. Install the sealing ring onto the front of the connector housing. This
seal is held in place when the connector is locked onto the
instrument. See Figure 3-11.
Figure 3-11 Install Connector Seal
17. Slide the large lock ring down the cable and over the cable housing
body. It may need a twisting action to fully position it against its
retainer ring. Verify that its larger opening is facing outward, away
from the cable. See Figure 3-12.
Figure 3-12 Cable Connector Completed
Note: It is presumed that the REMOTE-4PN Series will be used in a
network of RS485/MODBUS instruments. Therefore, references are
made to a "hub" or the "hub end" of the ethernet cable.
The "hub" may be a computer equipped with an RS232 adapter, an
RS485 hub or one of many RS485 data gathering equipment.