TruScan 532 Single - Wavelength Operators Manual Rev. No 02 Page 21 of 114
Section 3 PRODUCT SPECIFICATIONS
3.1 TruScan 532 Single-Wavelength Pattern Scanning Laser System
Specification
The following are the System Specifications for the TruScan 532 Single-Wavelength
Ophthalmic Laser.
Console Laser System
General Specification
Electrical
Input
100 to 230 Vac. 50/60 Hz Single phase
Power
400VA
Fuse rating
T3.15AH250V @ 100-230Vac (Time Lag)
Temperature
Range
Transport: -10 to 70°C
Operating: 15 to 30°C
Storage: -10 to 55°C
Relative
Humidity
Range
Operating: 30% - 85% non-condensing
Storage and Transport: up to 95% non-condensing
Atmospheric
pressure
Operating: 800-1060 mbar
Storage and Transport: 500-1060 mbar
Cooling
System
Fan cooled and TEC’s for Laser Diode and Crystal
Dimensions
130mm(H) x 370mm(W) x 330mm(D)
Weight
13 Kg (System) 20 Kg (Packed)
Treatment Laser
Laser Type
Diode Pumped Frequency
Doubled YAG
Wavelength
532nm
Mode of
Operation
CW
Power
Output
2 W Maximum to Patients
Cornea Power from Delivery
units laser aperture displayed
on LCD or Remote Display
Panel
Power
Adjustment
Variable from 0.05 to 2.0 W