Now place the Gap Setting Tool on the Reference
sensor pin (R) and turn the tool until the Refer-
ence Sensor indicator ‘Ref’ just turns green.
Then click Ref’. The Lid will now move down
500mm, so that the ‘Lid’ Sensor can be set.
Place the Gap Setting Tool on the Lid sensor pin
(L) and adjust until the Lid Indicator ‘Lid’ just
turns green. When set, click the ‘Lid’ button.
The motor will now move the Lid down 1000mm.