Operator & Service Manual
/ LiteGait®
Jr.
Jr.
Jr. / LiteGait®
MX
MX
MX LiteGait®
LiteGait® is a Registered Trademark of Mobility Research, Inc.

Questions and Answers

  • C
    Christopher OrtizAug 18, 2025
    How to fix LiteGait Medical Equipment actuator that doesn't raise or lower?
    • T
      Timothy Pierce IVAug 18, 2025
      If the actuator on your LiteGait Medical Equipment isn't raising or lowering, there could be a few reasons. First, check the battery charge and charge it fully following the charging instructions. If charging doesn't solve the problem, it might need replacement. Also, ensure that all cable connections to the control box are firmly connected. If these steps don't resolve the issue, it could be a power system problem.
  • J
    Jacob BrownAug 20, 2025
    What to do if LiteGait Walkable Medical Equipment battery does not charge?
    • A
      amber61Aug 20, 2025
      If your LiteGait Medical Equipment battery isn't charging, ensure the charge cable is connected to a working 110V power outlet (or 220V for international customers). Check that the charging cord is properly connected. The green LED should turn red when charging and green again when fully charged. Also, test the power outlet with another device to confirm it's working.
  • D
    Deanna EdwardsAug 23, 2025
    How to troubleshoot LiteGait Walkable BiSym display not powering on?
    • R
      rodriguezstephenAug 23, 2025
      If the BiSym display on your LiteGait Medical Equipment isn't powering on, first check that the battery cable is connected to the BiSym display. Also, ensure the BiSym battery is charged by following the charging procedure. The Digital BiSym is powered by a separate battery located near the top of the LiteGait.
  • D
    David AndersonAug 26, 2025
    What to do if LiteGait Walkable Medical Equipment BiSym display is not reading weight?
    • T
      townsendstephenAug 26, 2025
      If the BiSym display on your LiteGait Medical Equipment isn't reading weight, check the cables on the left side of the scale. There should be three cables: two phone jack-like connectors and one connecting to the top of the LiteGait. Make sure these are securely attached to the BiSym display.
  • S
    Shane DixonAug 28, 2025
    Why LiteGait Walkable device does not roll easily?
    • M
      Michelle SuarezAug 28, 2025
      If your LiteGait Medical Equipment device isn't rolling easily, check if the total locking casters, labeled with a red BRAKE sticker, are unlocked.
  • C
    Charles FoxAug 30, 2025
    What to do if LiteGait Walkable Medical Equipment BiSym display is not reading zero when no weight is on it?
    • R
      Robin PalmerAug 30, 2025
      If your LiteGait Medical Equipment BiSym display isn't reading zero when there's no weight on it, note that slight movement in the harness can cause some fluctuation. A reading near zero is normal.
  • J
    James FriedmanSep 1, 2025
    What to do if LiteGait Medical Equipment does not stay stationary when locked?
    • E
      Emily MurraySep 1, 2025
      To ensure your LiteGait Medical Equipment stays stationary when locked, make sure all four casters are locked.
  • J
    James BrightSep 4, 2025
    Why patient is complaining of groin or harness discomfort using LiteGait Walkable Medical Equipment?
    • A
      amywilliamsonSep 4, 2025
      If a patient is experiencing groin or harness discomfort while using the LiteGait Medical Equipment, ensure the harness wrap and groin piece are securely tightened. The harness wrap should be tight enough to grip the tissue around the abdomen, and the groin piece should prevent the harness from riding up. You can also add padding around the abdomen if needed.
  • V
    Victoria MercerSep 7, 2025
    How to prevent LiteGait Medical Equipment harness from riding up and causing pressure?
    • D
      Dennis CannonSep 7, 2025
      To prevent the LiteGait Medical Equipment harness from riding up and causing pressure in the groin area, securely tighten the bottom two straps on the harness wrap. Only tighten the top strap if it sits below the rib cage.
  • K
    kingdavidSep 9, 2025
    How to alleviate chest discomfort from LiteGait Walkable Medical Equipment overhead straps?
    • R
      Ronald PettySep 9, 2025
      If frontal overhead straps are causing discomfort in the chest area for female patients using LiteGait Medical Equipment, try using an extender to increase the front panel size and distance between the overhead straps. Alternatively, ensure the harness wrap is placed with the opening in the back.

Summary

Serial Number

Important Safety Instructions

LiteGait Assembly

LG WalkAble Assembly

LiteGait Diagram

About Your LiteGait

Using Your LiteGait

Yoke Adjustment

Manual adjustment of the LiteGait yoke height using knobs and spring pin.

FlexAble Functionality

Allows the yoke to become position flexible, offering up to 5 inches of travel for patient preference.

Raising and Lowering MX Yoke

Describes how to raise and lower the MX yoke using a hand switch with up/down arrows.

Integrated FreeDome Operation

Explains how to rotate and lock the FreeDome for patient rotation support.

Charging the LiteGait Battery

Instructions for removing and charging the LiteGait battery pack using the provided charger.

LiteGait Charging Procedure

Detailed steps for charging the LiteGait battery, including indicator lights and duration.

Adjusting Handlebars

How to loosen, adjust height, and retighten the handle bars for patient use.

Base and Casters Description

Description of the four casters (total locking and directional) and their function.

Starting Optional BiSym Bluetooth

Instructions for starting and using the optional BiSym Bluetooth system.

BiSym Patient Orientation

How to change patient orientation relative to the LiteGait device using the BiSym interface.

BiSym Display Units

How to display support values in weight or percentage units on the BiSym display.

BiSym Time Duration Setting

How to set the maximum duration for data collection sessions on the BiSym monitor.

BiSym Set User Weight

Instructions on setting the user's weight on the BiSym system for accurate readings.

BiSym Set Goal (Weight;Percentage)

How to set weight or percentage goals for patient support in the BiSym system.

BiSym Patient Feedback Display

Explanation of the patient feedback display showing support levels and goal achievement.

BiSym Tablet Holder Installation

Instructions for installing the tablet holder for the BiSym system.

Charging BiSym Battery and Display

Steps for charging the BiSym display's internal battery via USB.

Harness Application Overview

Describes the harness components and how they provide support during gait therapy.

Preparing Harness for Application

Steps to select, adjust, and attach the groin piece to the harness for proper fit.

Estimating Harness Starting Size

Method to estimate the correct harness girth size before patient application.

Harness Symmetry Test

Procedure to check and ensure symmetry of the harness buckles for even support.

Harness Application While Standing

Detailed steps for applying the harness to a patient while they are standing.

Attaching Groin Pieces

Instructions for connecting and tightening the groin piece straps for secure fit.

Harness Application in Supine

Steps for applying the harness to a patient who is lying down (supine).

Leg Strap Application

How to correctly wrap and attach the leg straps for optimal support and positioning.

Connect Harness to LiteGait

Steps for connecting the patient's harness to the LiteGait unit for support.

Harness Connection: Lifting Not Necessary

Alternative harness connection method for patients who don't need assistance to stand.

Stepping Onto Treadmill

Procedure for assisting a patient in stepping onto a treadmill using the LiteGait.

Stepping Onto Treadmill (Continued)

Continued steps for safely moving the LiteGait unit and patient onto a treadmill.

Over Ground Therapy Guidelines

Guidelines for using the LiteGait for over ground gait training and other activities.

Unit Care and Maintenance

Cleaning Frame, Grips, and Covers

Procedures for cleaning the frame, handle grips, and covers using approved solutions.

Cleaning the iHarness

Instructions for cleaning the iHarness wrap and overhead straps with disinfection wash.

Laundering Harness and Straps

Guidelines for washing and drying all harnesses and groin straps according to facility protocols.

Harness Storage Guidelines

Recommendations for proper storage of the harness to prevent damage to buckles and maintain effectiveness.

Unit Maintenance Checklist

A checklist for regular inspection of LiteGait components and recommended replacement schedules.

Maintenance Contact Information

Contact details for Mobility Research Service & Parts Department for maintenance inquiries.

Buckle Assembly Checks

Checks for tightness and secure attachment of buckle assemblies on yoke arms.

Battery Replacement Guidance

Guidance on when to replace a battery that loses charge quickly or requires frequent charging.

Power System Checks

Testing the actuator's UP/DOWN function and identifying potential battery issues.

Knobs and Pins Check

Ensuring knobs screw easily and secure the yoke and handlebars properly when tightened.

Total Locking Casters Functionality

Verifying casters lock in place, preventing swivel and wheel rotation when brake is pressed.

Directional Locking Casters Functionality

Checking that steer tab locks swivel when aligned with the base leg, allowing forward/backward movement.

Troubleshooting

Actuator Not Raising or Lowering

Diagnosing issues with the actuator not raising or lowering, covering battery and connections.

Battery Not Charging

Troubleshooting steps for when the LiteGait battery fails to charge, checking connections and outlet.

Battery Not Holding Charge

Addressing issues where the battery does not hold a charge, suggesting charging or replacement.

Device Does Not Roll Easily

Identifying causes for the device not rolling easily, such as locked casters or misalignment.

Device Does Not Stay Stationary When Locked

Resolving issues where the device moves when casters are locked, checking for proper locking.

Patient Groin or Harness Discomfort

Addressing patient discomfort by ensuring proper harness and groin piece tightness and fit.

Harness Riding Up; Groin Pressure

Resolving harness riding up by ensuring secure tightening of bottom straps and proper fit.

Overhead Straps Cause Chest Discomfort

Solutions for chest discomfort from overhead straps, potentially using extenders.

Overhead Straps Slipping Off Shoulders

Addressing slipping overhead straps by adjusting harness position or opening.

Patient Cannot Stand for Harness Adjustment

Alternative harness application in supine for weak patients, avoiding sitting.

BiSym Display Not Powering On

Troubleshooting steps for a BiSym display that does not power on, checking battery and connections.

BiSym Display Not Reading Weight

Diagnosing why the BiSym display is not reading weight, checking load cell cable connections.

BiSym Display Not Reading Zero

Addressing BiSym readings not showing zero, checking harness movement or calibrating scale.

BiSym Auto Power Down Issue

Resolving BiSym auto shut-off issues by checking or adjusting settings.

Parts List

Parts List: Harness and Accessories

Lists various harness components like wraps, groin pieces, and straps with part numbers.

Parts List: Power System (LG100 MX)

Lists power system components including actuator, battery, control box, and charger for LG100MX model.

Parts List: Power System (LG200 MX)

Lists power system components for the LG200MX model.

Parts List: Base Components

Details various base frame components and sizes with corresponding part numbers.

Parts List: Handlebars (LG75;LG100 W;LG100 MX)

Lists handlebar components for specific LiteGait models.

Parts List: Handlebars (LG150;LG200;LG200 MX)

Lists handlebar components for LG150/LG200/LG200MX models.

Parts List: Yoke Components

Lists complete yoke assemblies, socks, and cartridges with part numbers and descriptions.

Parts List: Post Components

Details post components including complete posts, caps, and screws with part numbers.

Parts List: Digital BiSym Components

Lists components for the Digital BiSym system, including scale, battery, and charger.

Resource Directory

LiteGait Walkable Specifications

General IconGeneral
BrandLiteGait
ModelWalkable
CategoryMedical Equipment
LanguageEnglish