Operator Manual Valiant
7
· Operation in close proximity (i.e. less than 2 meters) to working short-wave or
microwave therapy units may produce instability in the Valiant output.
· To prevent electromagnetic interference, we strongly advise using separate mains
groups (phases) for the Valiant and the short-wave or microwave equipment, and a
distance of at least 2 meters is kept between the Valiant and the short-wave or
microwave equipment.
· Ensure that the mains cable of the short-wave/microwave unit does not come near the
Valiant or the patient.
If problems with electromagnetic interference persist, please contact your authorized
distributor.