Operator Manual Valiant
19
7 Technical specifications
Supply-voltage : 230 V, +/- 10%, 50-60 Hz. *
Driving-force : Frequency-controlled rotary-field electromotor.
Motor-capacity : 1,9 kW
Max. power-consumption : 2500 VA
Min./Max. speed
Valiant : 0.5 – 20 km/h
Valiant Rehab : 0.1 – 12 km/h
* The Mains supply should be 230 volt. When less, e.g. 110 volt, the Valiant should be connected between two phases.
Belt area : 50 x 150 cm (eff.)
Max. permissible weight on belt : 160 kg
Inclination : 0-25%
Dimensions (LxWxH in cm) : 213 x 90 x 135
Weight : 149 kg
Classification:
Medical class Valiant : IIa type B, IEC 601-1 chapter 18, 19 and 20
Safety class 1
The equipment has a safety earth (ground) connection and must be connected to an earthed
(grounded) wall socket.
Type B
The leakage current and safety earth correspond to standard IEC 60601-1
Peripheral port:
RS-232 PC port :19200 Baud, 8 data bits, no parity, 1 stop bit. (See Chapter 4.9.) Extra
devices (computers, e.g.) which are connected to the Valiant, have to be placed at least 2
meters from the Valiant.
The extra devices need to be according to the local safety regulations.
The Mains connector of the Valiant and extra devices need to be according to the local safety
regulations for medical devices.
7.1 Environmental conditions
Permissible
conditions
during:
Temprature
[min - max ºC (ºF)]
Relative humidity
[%] non condensing
Use
5 – 40 (41 - 104)
30 – 90
Transport
-20 – 60 (-4 - 140)
10 – 95
Storage
-20 – 60 (-4 - 140)
10 – 95
Technical modifications reserved.