Model Numbers
UTRIX-16 - Multi-Trip PDF Temperature Recorder with dual alarms
UTRIX-16M - Multi Trip PDF Temperature Recorder with 6 alarms
Temperature Sensor Measurement Range
-25°C to +7C (-13°F to +158°F)
Operating Temperature Range
-25°C to +7C (-13°F to +158°F)
Storage Temperature Range
0°C to +40°C (3F to +104°F)
Ambient humidity range during transport and use
0 to 95%RH
Sensor Type
Precision thermistor
Rated Temperature Resolution
0.1°C (0.1°F) across entire range
Rated Temperature Accuracy
l Better than ±0.5°C (±0.9°F) for -5°C to +30°C (-23°F to 86°F)
l Better than ±0.7°C (±1.5°F) for -25°C to -5°C (-13°F to 23°F)
l Better than ±0.7°C (±1.5°F) for +40°C to +60°C (86°F to 140°F)
l Better than ±0.8°C (±1.5°F) for +60°C to +70°C (86°F to 140°F)
Sensor Reaction Time
Typically less than 7 minutes (T90) in moving air (1m/s), method as detailed
in EN12830:1999.
Clock accuracy
Quartz crystal locked real time clock, rated accuracy ±25ppm @ 25°C (equiv to
2.5 seconds/day)
Rated temperature coefficient is -0.030.006ppm/°C (i.e. typically +/-
0.00294seconds/day/°C)
Recording Capacity
16,129 real time temperature values, giving
l 112 days @ 10min logging or
l 168 days @ 15min logging.
Memory type
Non volatile
Sampling Interval
Configurable from 30 seconds to 18 hours.
Start options
l Push button start with optional configurable start delay from 1
minute to 72 hours
l Date/time start
Alarm functions
l one configurable upper alarm
l one configurable lower alarm
l four additional, freely configurable alarms (UTRIX-16M only)
l Flashing OK and Alert indicators, linked to alarms
Vibration
Withstands vibration specification as detailed in EN12830:1999
Shock
l Withstands shock specification as detailed in EN12830:1999
l Withstands 5 drops from 1m onto smooth concrete floor without loss
of function or calibration
EMC compliance
l EC EMC directives (EN 61000-6-1:2005 & EN 61000-6-3:2006)
l Includes electrostatic discharge as prescribed in EN 61000-4-2
l Complies with FCC Part 15 Subparts A and B
Environmental
IEC 60529: IP64 with USB cap fitted
Fits into IP67 Protective Enclosure 200-000020
Case Material
Polycarbonate
Power source
CR2032 3V Li-MnO
2
battery non user-replaceable, non-rechargeable
Battery life
Shelf life of up to two years before configuration.
Typically 2 years of operation with normal use (6 minute logging, download
data monthly).
Size
93mm(H) x 54.5mm (W) x 8.6mm (T) including USB cap
Weight
39g
Calibration
Calibration traceable to an ISO/IEC 17025 accredited testing laboratory
PDF features
l compliant with standard 1.6 and later
l Single page report with trip and alarm summary
l Multi page report with list of readings including date/time
Download time
l Typically with full memory (16,129 readings) less than 30 seconds
from time of insertion to availability of PDF report.
l Typically less than 10 seconds from time of insertion to availability
of LTD file in LogTag
®
Analyzer
Software requirements
l LogTag
®
Analyzer version 2.5 or later to configure and download
(2.8 or later for UTRIX-16M)
l PDF reader software to access onboard PDF files
USB compatibility
USB 2.0, type A plug
Accessories
l Wall holder 200-000010
l IP67 Protective Enclosure 200-000020
l Replacement protective cap 200-000425