ISR 12-LO IGAR 12-LO Operating Manual Troubleshooting 41
9 Troubleshooting
Before sending the pyrometer for repair, try to find the error and to solve the problem with the
help of the following list.
Temperature indication too low
Incorrect alignment of the pyrometer to the object.
New correct alignment to achieve the max. temperature signal (see 3.6).
Incorrect focusing of the optics.
Focus correctly to the measuring distance if necessary change lens (see 3.5).
Emissivity slope set too high.
Set lower correct emissivity slope corresponding to the material (see 4.6).
Temperature indication too high
Emissivity slope set too low.
Set lower correct emissivity slope corresponding to the material (see 4.6).
The measurement is influenced by reflections of hot machine parts.
Use mechanical construction to avoid the influence of the interfering radiation
(sighting tube).
Measuring errors
Indicated temperature is immediately 1 °C below beginning of temperature range.
Contamination of the lens or broken fiber.
Clean lens or control the fiber. Recommendation: use of air purge (see 2.6).
Indicated temperature is immediately 1 °C below beginning of temperature range, although
the air purge unit is used. Probably compressed air is not clean or air failed.
Clean the lens and use clean, dry and oil free compressed air.
HF-interferences.
Correct the connection of the cable shield (see 4).