LS200-INS-LAB-RevB17 14
Lowpressurealarm
ALARM RESET Button
Pump Front Panel
The pump is equipped with an audible low pressure alarm which enables the pump to audibly
alert the caregiver to a low pressure issue with the mattress system.
The low pressure alarm can be muted for twenty minutes by pressing the pump front panel
ALARM RESET button, shown above.
WARNING:Thelowpressurealarmrequiresoperatorresponsetopreventthepatient
frombottomingout.
Info: The low pressure alarm is automatically disengaged for 45 minutes when control
unit is first powered on to ensure no alarms are raised during initial inflation.
CPR DEFLATION
WARNING:Forquickmattressdeation:
1. Levelthebed.
2. Turnthepumpoffanddisconnectthehoseconnectorfromthepump.
3. Lowerorremovesiderailoncaregiver’ssideifnecessary.
4. ReleasetheCPRquickdeationvalveatheadendofthemattress(patientleft).
AfterCPRisperformed
1. Reconnect pump hoses to mattress and turn pump back on.
2. Raise or install side rail if necessary.
3. Recongure bed and accessories as in initial placement.