7 CARE AND MAINTENANCE
Proper care and maintenance are essential to keeping your Select Mattress System in a safe
operating condition. In addition to inspecting the unit before each use, periodic maintenance
checks should be done.
WARNING:Inspectthemattressbeforeeachuse.Ensureallhardwareandaccessories
aresecureandthepumpisfunctioningproperly.
WARNING:ServiceandrepairoftheSelectMattressSystemMUSTbeperformedby
qualiedpersonnelONLY.
WARNING:UnauthorizedmodicationoftheMattressSystemortheuseofnon-Lumex
replacementpartsmaychangethestructureoftheMattressSystemandcouldcreatea
hazardouscondition,whichmayresultinseriousinjuryandwillvoidthewarranty.
WARNING:Thepumpcontainsnoserviceablecomponents.DONOTattempttoopen
thepump.Ifserviceisrequired,consultGFTechSupportat1-770-368-4700forfurther
information.
When you believe a component or part is not functioning properly, immediately contact GF
Tech Support at 1-770-368-4700, as a potentially hazardous condition could exist.
CLEANING
CleaningtheMattress
Surface wipe weekly with neutral detergent and lukewarm water. Rinse thoroughly with water
and dry thoroughly.
CleaningtheCover
If the top or bottom cover becomes severely soiled remove it, clean as follows, and replace
with a clean cover. Covers can be washed and thermally disinfected in a washing machine.
s NOTICE:Neverusephenolbasedcleaningsolutions.
Industrial cleaning
Break washes cold 10 minutes
Main wash 60°C 6 minutes
Main wash 72°C 10 minutes
Extraction 2 minutes
3 cold rinses
Extraction 5 minutes
Domestic cleaning
Pre-wash cold
Main wash 72°C 10 minutes
Extraction 2 minutes
Cold rinses
Extraction 5 minutes
Tumble Drying or Tunnel Drying is not recommended.
Mattress Cells can be wiped over with a solution of sodium hypochlorite 1000
ppm or any other non-phenolic germicidal solution.