LS200-INS-LAB-RevB17 16
CleaningthePump
s NOTICE:Donotoodanypartofthepumpwithcleaningsolution.Donotimmersethe
pumpinliquid.
s NOTICE:Avoidspillinganyliquidonpump.Ifspillsdooccur,cleanuidfrompump
wearingrubberglovesorwhileunitisunpluggedtoavoidanypossibilityofshock.
Onceliquidisremoved,checkoperationofcomponentsinareaofspill.Anyliquid
remainingonpumpcancausecorrosion,whichmaycausecomponentstofailor
operateerratically,possiblyproducingpotentialhazardsforpatientandstaff.
Clean the pump weekly using a clean, damp soft cloth and mild detergent.
The pump casing is manufactured from ABS plastic and if the case is soiled the pump can be
wiped down with a sodium hypochlorite solution to dilution of 1000 ppm or any EPA-approved
hospital grade disinfectant (do not use phenol base cleaning solution).
INSPECTION / SYSTEM CHECK-OUT
Check each of the following before placing the Mattress System with a new patient:
1. Check mattress surface for tears or cracking; do not use if tears or cracks are present.
2. Ensure mattress is free of stains and is not overly faded.
3. Ensure air inlet hoses and connectors on mattress and pump are clean and undamaged.
4. Ensure pump and power cord are clean and undamaged.
5. Ensure pump hanger brackets are secure and operate correctly.
6. Ensure ON / OFF Power switch and comfort control knob both operate correctly.
7. Attach pump to Alternating Pressure hoses and power up to ensure there are no air leaks.
MAINTENANCE
Replaceairlter
Air Filter
Cover
1. Remove air lter cover (shown above).
2. Replace used lter with new lter.
3. Replace air lter cover.