Lamp Countdown Sequence
The system counts down the number of days unl a lamp change is required.
BLACKCOMB
6.1
BLACKCOMB
5.1
BLACKCOMB
4.1
At thirty days remaining, the LED or display screen will change to a yellow cauon indicator. At
seven days remaining, the system will addionally repeat an audible chirp. Past the zero day
threshold, the LED or display screen changes to solid red with a connuous buzzer.
BLACKCOMB
6.1
BLACKCOMB
5.1
BLACKCOMB
4.1
At any point during this sequence, the audible chirp or alarm can be deferred for seven days
by holding the controller buon down for a period of ve seconds. The number of deferrals
used will be displayed as below. Once the deferral expires, the alarm will sound once again. The
deferral can be repeated up to three mes. PLEASE NOTE: At any point aer lamp expiraon,
the water may be unsafe for consumpon and should not be consumed without another form
of disinfecon.