EasyManua.ls Logo

Lutron Electronics CD-4318SD - Page 12

Lutron Electronics CD-4318SD
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3) Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
4) Hold the " Probe Handle " ( 3-18, Fig. 1) by hand
and let the " Sensing head " ( 3-19, Fig. 1 )
immersed wholly into the measured solution. Shake
the probe to let the probe's internal air bubble drift
out from the sensing head.
Display will show the conductivity mS ( uS ) values.
5) Use the two fingers to press the " Button " ( 3-5, Fig. 1 ) " ,
and " Button " ( 3-6, Fig. 1 ) at the same
time. the display will show the following screen as
example(Fig. a), release the both fingers , then Press the
" Enter Button " ( 3-8, Fig. 1 ) , the display will show
the following screen as example (Fig. b).
CD
CAL Fig. a CAL Fig. b
6) Press the " Enter Button " ( 3-8, Fig. 1 ), the measuring
value will present on both upper and lower Display.
Use " Button " ( 3-5, Fig. 1 ), " Button " ( 3-6, Fig. 1 )
to adjust the up display value exact same as the standard
conductivity value. Press the " Enter Button " ( 3-8. Fig. 1 )will
save the calibration data and finish the calibration procedures.
Standard value
1.318 1.413
1.318 1.318
*
If only intend to make the one point calibration,
just execute the 2 mS range ( 1.413 mS Cal. )
is enough.
*
Multi-points calibration procedures should
execute the 2 mS range ( 1.413 mS Cal. )
calibration at first, then make other ranges
(20 uS range, 20 mS range or 200 mS range )
calibration procedures following if necessary.
10

Related product manuals