10. ADVANCED SETTING
Under do not execute the Datalogger function,
press the " SET Button " ( 3-7, Fig. 1 ) continuously at
least two seconds will enter the " Advanced Setting " mode.
then press the " SET Button " ( 3-7, Fig. 1 ) once a while
in sequence to select the eight main function, the
display will show :
DAT
E.....
.................
Set clock time ( Year/Month/Date, Hour/Minute/
Second )
SP T
......
.......................
Set sampling time ( Hour/Minute/Second )
PO
FF......
....................
Auto power OFF management
BEEP.......................
Set beeper sound ON/OFF
DEC.........................
Set SD card Decimal character
SD
F.....
..............
SD memory card Format
T CF......................
Select the Temp. unit to or
PCNT
......
............
Set CD temperature compensation factor
CELL
......
...........
CELL K Value setting
Remark :
During execute the " Advanced Setting " function,
if press " HOLD Button " ( 3-3, Fig. 1 ) will exit the
" Advanced Setting " function, the LCD will return
to normal screen.
10-1 Set clock time ( Year/Month/Date,
Hour/Minute/ Second )
When the upper display show " DATE "
1) Use the " Button " ( 3-5, Fig. 1 ) or " Button "
( 3-6, Fig. 1 ) to adjust the value ( Setting start from
Year value ). After the desired value is set, press the
" Enter Button " ( 3-8, Fig. 1 ) once will going to
next value adjustment ( for example, first setting
value is Year then next to adjust Month, Date, Hour,
Minute, Second value ).
20
Remark :

Related product manuals