EasyManua.ls Logo

Lutron Electronics WAC-2019SD - Page 17

Lutron Electronics WAC-2019SD
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-3 mV Measurement
The instrument build in mV ( millivolt ) measurement
function, which enable you to make ion-selective, ORP
(oxidation-reduction potential), and other precise mV
measurements.
1) Under the meter is operated under the " PH " Mode.
Press the " Function Button " ( 3-6, Fig. 1 ) once,
the measure unit " pH " will change to " mV "
* Press the : Function Button " once will return to
" pH " function.
2) Prepare the ORP Electrode ( optional, ORP-14 ), install
the " Probe Plug " of ORP electrode into the " PH
Socket/BNC Socket " ( 3-16, Fig. 1 ).
3) The Display will show the mV value.
5-4 pH calibration
Calibration Consideration
The most ideal pH ELECTRODE generates 0 mV at pH
7.00 ( 177.4 mV at PH 4 ) and meter has been always
calibrated with signals which simulate the most ideal pH
ELECTRODE ( based on 25 ambient environment ).
However not every pH ELECTRODE is as accurate as the
most ideal one, so calibration procedures are necessary
to be done before the first time measurement.
In addition to the first time measurement, users are also
recommended to execute the calibration procedures to
ensure the high accuracy measurement.
15

Related product manuals