EasyManua.ls Logo

Lutron Electronics WAC-2019SD - Page 24

Lutron Electronics WAC-2019SD
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3) Power on the meter by pressing " Power Button "( 3-2, Fig. 1 )
> 2 Sec., press the "Function (3-6 Fig1)" Select the Meter's
measurement screen to the only " Conductivity "screen.
4) Hold the " Probe Handle " ( 6-2, Fig. 3 ) by hand and let the
" Sensing head " ( 6-3, Fig. 3 ) immersed wholly into the measured
solution. Shake the probe to let the probe's internal air bubble drift
out from the sensing head.Display will show the conductivity mS
( uS ) values.
5)
Use the two fingers to press the "
Button " ( 3-5, Fig. 1 ),
and "
Button " ( 3-6, Fig. 1 ) ,Until Display will show the
following screen then release the both fingers.
Temp. Calibration
Calibration
=>
Zero Calibration
Gain Calibration
Clear
Escape Calibration
Press the "
Button " ( 3-5, Fig. 1 ) or "
Button " ( 3-6, Fig. 1 )
select to Gain calibration point.
6) Press the " Enter Button " ( 3-8, Fig. 1 ), the measuring
value will present on both upper and lower Display. Use
"
Button " ( 3-5, Fig. 1 ), "
Button " ( 3-6, Fig. 1 ) to adjust
the up display value exact same as the standard conductivity
value. Press the " Enter Button " ( 3-8. Fig. 1 ) will save the
calibration data and finish the calibration procedures.
Standard value
* If only intend to make the one point calibration, just
execute the 2 mS range ( 1.413 mS Cal. ) is enough.
22

Related product manuals